8 062 operational procedure for particle count
| Standard Operating Procedure | SOP No. |
|---|---|
| Operational Procedure for Particle Counts-ISO 14644-1 | 8.062 |
| Effective:08-18-09 Supersedes:08-17-09 Revision Date:08-18-09 Page No.:1of5 | |
| DCR No.:09004 Revision No.:1 |
1
Purpose
1.1 To establish a standard procedure for verifying airborne particle counts in the static (empty shell): as-built, at-rest, or operating facility modes is within the limits of Class concentrations of airborne particles per ISO 14644-1 standard.
2
Scope
2.1 This procedure applies to the airborne optical particle counters, which are used to measure the actual particle count level, temperature and relative humidity.
3
References
3.1 ISO 14644-1: Cleanrooms and Associated Controlled Environments – Classification of air cleanliness, 1999.
3.2 NEBB Procedural Standards for Certified Testing of Cleanrooms, Second Edition, 1996.
4
Definition
4.1 As-Built Facility
4.1.1 A cleanroom that is complete and operating, with all services connected and functioning, but has no production equipment or operating personnel within the facility.
4.2 At-Rest Facility
4.2.1 A cleanroom that is complete and has the production equipment installed, but has no personnel within the facility.
4.3 Operating Facility
4.3.1 A cleanroom in normal operation.
5 Responsibility
5.1 TAB technicians shall record all test readings on Form FN 8.016.1
5.2 TAB technicians shall save all test reports in files located in the TAB Department of Therma.
5.3 All test equipment utilized shall be in calibration in accordance with NEBB Institute of Standards and traceable to the National Institute of Standards and Technology (NIST).
6 Materials Requirement
6.1 None
7 Test Equipment
7.1 Airborne Optical Particle Counter
8 Procedures
8.1 Verify that all aspects of the cleanroom system which contribute to its operation integrity (air handling, filtration systems, walls, ceilings, floors, etc.) are complete and functioning. Verify room pressure relationship differentials are correct.
8.2 Determine the class of the tested room or work zone area from the supplied documentation.
8.3 Using ISO 14644-1 procedure the appropriate number of saten, divide the room or work zone into sampling location points, no less than
8.4 Identify the location of the sampling points.
8.5 Use the particle counter to measure the particle count at each location at the specified particle sizes which range from 0.1 $ \mu\mathrm{m} $ to 5.0 $ \mu\mathrm{m}. $ (Refer to user manual for correct operation of particle counter.)
8.6 Record the particle count, temperature, and relative humidity at each location when the readings are stabilized.
8.7 Calculate the mean particle count at each grid location and record on the Form FN 8.016.1 (Particle Count Readings) and FN 8.016.2 (Certification of Cleanroom Compliance).
8.8 Check the mean values of the particle count with Diagram A.
8.9 Record any specified environmental conditions at the time of testing.
9 Review and Approval
9.1 Return the Forms FN 8.016.1 (Particle Count Readings) and FN 8.016.2 (Certification of Cleanroom Compliance) to the TAB Department for review.
Diagram A
| ISO classification number(N) | Acceptance Criteria Class Limits(particles/m3of air) | |||||
|---|---|---|---|---|---|---|
| ISO classification number(N) | 0.1μm | 0.2μm | 0.3μm | 0.5μm | 1.0μm | 5.0μm |
| ISO Class 1 | 10 | 2 | ||||
| ISO Class 2 | 100 | 24 | 10 | 4 | ||
| ISO Class 3 | 1000 | 237 | 102 | 35 | 8 | |
| ISO Class 4 | 10000 | 2370 | 1020 | 352 | 83 | |
| ISO Class 5 | 100000 | 23700 | 10200 | 3520 | 832 | 29 |
| ISO Class 6 | 1000000 | 237000 | 102000 | 35200 | 8320 | 293 |
| ISO Class 7 | 352000 | 83200 | 2930 | |||
| ISO Class 8 | 3520000 | 832000 | 29300 | |||
| ISO Class 9 | 35200000 | 8320000 | 293000 |